本文从三个方面介绍了721型分光光度计波长误差的调整: (1)721型分光光度计的单色器中,因选择波长的转动螺线轮设计较好,波长度盘示值间隔基本均匀,故当波长误差在几个波长区皆正、皆负且数值接近时,仅松动度盘的三个紧固螺钉,转动波长度盘,即可调节成功。 (2)当721型分光光度计长短各区的波长误差数值不一致,但差别不大时,旋转仪器左侧准直镜的调节螺钉,调节其沿铅垂轴的转角方位,使各点波长示值误差符号相同和数值相近后,再转动波长盘也可调节成功。 (3)当以上两种方法仍不能使波长示值合格时,也就是在仪器波长误差较大时,首先,应注意记录短、长两波长检定点的示值差
This paper introduces the adjustment of wavelength error of 721 spectrophotometer from three aspects: (1) In the monochromator of 721 spectrophotometer, due to the design of the rotating spiral wheel with the selected wavelength is good, Even when the wavelength error is in a few wavelength areas are positive and negative, and the numerical value is close, only the three fastening screws of the dial are loosened, and the rotating dial can be adjusted successfully. (2) When the wavelength error of the 721 spectrophotometer is inconsistent, but the difference is not very big, rotate the adjustment screw of the collimator on the left side of the instrument to adjust the azimuth of its rotation along the vertical axis so that the wavelength value of each point shows After the error symbols are the same and the values are similar, then the wavelength plate can be adjusted and adjusted successfully. (3) When the above two methods still can not pass the wavelength indication, that is, when the wavelength error of the instrument is large, firstly, note the indication difference of the short and long two-wavelength verification points