ABL30血气分析仪阀阻值变大引起的RINSE ERROR故障检修

来源 :医疗装备 | 被引量 : 0次 | 上传用户:lsssml1990
下载到本地 , 更方便阅读
声明 : 本文档内容版权归属内容提供方 , 如果您对本文有版权争议 , 可与客服联系进行内容授权或下架
论文部分内容阅读
故障现象:开机后出现:RINSEERROR,工作无法进行。故障原理及分析:“RINSEERROR”即VSB传感器感受不到冲洗液的信号,也就是说冲洗通路堵塞或不畅,使VSB传感器检测不到冲洗液的信号。造成该故障原因有三点:一是由于泵管及软管阀V1、V5、V Symptoms: After starting up: RINSEERROR, work can not be carried out. Failure principle and analysis: “RINSEERROR” means that the VSB sensor does not sense the signal of the rinsing fluid, which means that the rinsing path is blocked or unfavorable, so that the VSB sensor cannot detect the rinsing fluid signal. There are three reasons for this failure: First, due to the pump tube and hose valves V1, V5, V