论文部分内容阅读
A benzoylferrocene modified multi‐wall carbon nanotube paste electrode for the measurement of methionine (MET) concentration is described. MET electrochemical response characteristics of the modified electrode in a phosphate buffer solution of pH 7.0 were investigated by cyclic voltammetry, square wave voltammetry, and chronoamperometry. Under optimized conditions, the square wave voltammetric peak current of MET increased linearly with MET concentration in the range of 1.0×107 to 2.0×104 mol/L. The detection limit was 58.0 nmol/L MET. The diffusion coefficient (D=5.62×106cm2/s) and electron transfer coefficient (α=0.4) for MET oxidation were also determined. The sensor was successfully applied for the measurement of MET concentration in human urine.
A benzoylferrocene modified multi-wall carbon nanotube paste electrode for the measurement of methionine (MET) concentration is described. MET electrochemical response characteristics of the modified electrode in a phosphate buffer solution of pH 7.0 were investigated by cyclic voltammetry, square wave voltammetry, and chronoamperometry . Under optimized conditions, the square wave voltammetric peak current of MET increased linearly with MET concentration in the range of 1.0 × 107 to 2.0 × 104 mol / L. The detection limit was 58.0 nmol / L MET. The diffusion coefficient (D = 5.62 × 106 cm 2 / s) and electron transfer coefficient (α = 0.4) for MET oxidation were also determined. The sensor was successfully applied for the measurement of MET concentration in human urine.