XED-150L X光机多故障检修三例

来源 :医疗设备信息 | 被引量 : 0次 | 上传用户:XIAO13075674309
下载到本地 , 更方便阅读
声明 : 本文档内容版权归属内容提供方 , 如果您对本文有版权争议 , 可与客服联系进行内容授权或下架
论文部分内容阅读
我科的日本岛津公司XED—150LX光机,因工作需要,从住院部搬迁到了门诊,但在搬迁后却发生了多起故障,现叙述如下:故障现象之一,机器在工作中,突然F-KV表针甩头,保护电路OVER灯亮,关机几分钟后再开机,机器又恢复正常,但工作一段时间后,故障又出现了。故障检查,根据 The Shimadzu XED-150LX light machine of our department was relocated from the inpatient department to the outpatient department because of work needs. However, after the relocation, many malfunctions occurred. These are described as follows: One of the symptoms is that the machine is suddenly working. F-KV hands shaking head, the protection circuit OVER light, turn off a few minutes after the boot, the machine has returned to normal, but after working for a period of time, the fault has emerged. Fault check, according to