Measurement of frequency stability in tunable lasers by using an F-P interferometer

来源 :Optoelectronics Letters | 被引量 : 0次 | 上传用户:wind2121
下载到本地 , 更方便阅读
声明 : 本文档内容版权归属内容提供方 , 如果您对本文有版权争议 , 可与客服联系进行内容授权或下架
论文部分内容阅读
A method for measuring the frequency stability of tunable laser is proposed by using confocal Fabry-Perot(F-P)interferometer.The F-P interferometer is used to get the output frequency of the laser as a reference,and the method eliminates the need of an independent optical source as a frequency reference.Using this technique,the frequency stability of the tunable external-cavity diode laser(ECDL)is measured to be 2.26×10~(-9)with an integration time of 20 ms. A method for measuring the frequency stability of tunable laser is proposed by using confocal Fabry-Perot (FP) interferometer is used to get the output frequency of the laser as a reference, and the method eliminates the need of an independent optical source as a frequency reference. Using this technique, the frequency stability of the tunable external-cavity diode laser (ECDL) is measured to be 2.26 × 10 -9 with an integration time of 20 ms.