目的提高~(99m)Tc-MIBI乳腺亲肿瘤显像的对乳腺癌的检出率.方法 采用我们设计的乳腺显像装置作~(99m)Tc-MIBI乳腺亲肿瘤显像,采用特殊俯卧位检查.其中108例的结果与病理资料对照.结果恶性病变组ECT阳性43/47;阴性4/47良性病变组ECT阴性56/61阳性5/61~(99m)Tc-MIBI乳腺亲肿瘤显像对乳腺恶性病变诊断的~(99m)Tc-MIBI乳腺亲肿瘤显像对乳腺恶性病变诊断的灵敏度91.48%;特异性为91.80%,诊断正确率为91.66%;假阴性率8.52%,假阳性率8.20%,阳性预测值89.58%,阴性预测值93.33%结论采用特殊显像装置并俯卧位检查,提高整个检查的准确性和灵敏度起了很大的作用,值得推广使用.
Objective To improve the detection rate of ~(99m)Tc-MIBI mammary tumour imaging for breast cancer. Methods We used a breast imaging device designed for ~(99m)Tc-MIBI mammary tumor imaging with special prone position. Examination. The results of 108 cases were compared with the pathological data. The results of malignant lesions were ECT positive 43/47; negative 4/47 benign lesions were ECT negative 56/61 positive 5/61~(99m)Tc-MIBI breast tumor imaging The diagnostic sensitivity of ~(99m)Tc-MIBI mammary gland tumor imaging for malignant breast lesions was 91.48% for the diagnosis of breast malignant lesions; the specificity was 91.80%, the diagnostic accuracy rate was 91.66%; the false negative rate was 8.52%, false positive rate. 8.20%, positive predictive value of 89.58%, negative predictive value of 93.33% Conclusion The use of special imaging devices and prone position examination to improve the accuracy and sensitivity of the entire inspection has played a significant role, it is worth promoting.