论文部分内容阅读
贝朗C型人工肾血透机特殊故障检修一例烟台毓璜顶医院杨凯,李首峰故障现象:在每次超滤(UF)测量结束时,均会出现电导度报警,大约5秒钟左右机器又恢复正常。分析检修:打开机器将A、B两液的泵管拆下,发现磨损较为严重,将其换新。再将机器的维修测试板插入功能...
A case of special troubleshooting of the Braun C type artificial kidney hemodialysis machine Yantai Yuding Hospital Yang Kai, Li Shoufeng Symptoms: At the end of each ultrafiltration (UF) measurement, electrical conductivity alarms will appear. The machine will recover in about 5 seconds. normal. Analysis and maintenance: Open the machine to remove the A, B two-fluid pump tube and find that the wear is serious and replace it with new one. Then insert the machine’s maintenance test board into the function...