论文部分内容阅读
应用102G气相色谱仪进行氟里昂二元混合物全浓度范围成份分析。用多点标样法测定相对摩尔校正因子,并按校正的面积归一化法定量计算。经五个二元混合物的测定结果证明,摩尔分数的绝对平均偏差为0.005左右。
The 102G gas chromatograph was used to analyze the full range of Freon binary mixtures. The relative standard molar correction factor was determined by the multi-point standard method and was normalized according to the corrected area. The results of the determination of five binary mixtures proved that the absolute mean deviation of the mole fraction is about 0.005.